Details of What's New in WinLTP 0.94

 

1.  Online measurment of Rs from the unfiltered trace, and the synaptic Peak Amplitude from the filtered trace

The best way to measure Rs and PkAmp online. Rs measurement was made from the capacitative peak transient from the unfiltered trace so as to not attenuate the transient peak (20 KHz sampling, 10 KHz external filtering). The PkAmp measurement of the synaptic currents was made from filtered trace so as to not measure the superimposed excess noise (20 KHz sampling, 10 KHz external filtering, 500 Hz internal filtering). The Rs and Rm measurements were made from the gray 10 KHz unfiltered trace, whereas the EPSCs would be measured from the blue internal 500 Hz filtered trace.

 